Management Representative versus PRRC: who does what?

For SaMD manufacturers, the difference between the Management Representative and the Person Responsible for Regulatory Compliance (PRRC) may not always be clear. While both roles are in place to ensure patient safety and regulatory compliance, they derive from different requirements.

What is the role of the Management Representative?

The Management Representative, also known colloquially as the Man Rep is a requirement from ISO 13485:2016 under §5.5.2:

“Top management shall appoint a member of management who, irrespective of other responsibilities, has responsibility and authority that includes:

  1. Ensuring that processes needed for the quality management system are documented:

  2. Reporting to top management on the effectiveness of the quality management system and any need for improvement;

  3. Ensuring the promotion of awareness of applicable regulatory requirements and quality management system requirements throughout the organization”.

Since February 2026, this is also aligned with the  FDA 21 CFR 820 Quality Management System Regulation (QMSR). The Man Rep reports to the Top Management and oversees the QMS, ensuring processes are well established and maintained. 

Who is Top Management? First things first, who is Top Management after all.  The ISO 13485 practical guide gives a very good explanation - essentially they are senior management individuals who can authorise, make decisions, responsible for ensuring resources are adequately provided to establish and maintain your QMS, and are ultimately responsible for the quality of your medical device. 

Your Quality Manual must depict who forms Top Management.  They can oversee the entire company, or just a sub-section of a wider organisation, in cases where the QMS is only covering a specific area of the organisation. 

Who can be appointed as the Man Rep?

The ISO 13485 practical guide again is helpful to provide additional clarity on the role of the Man Rep. The Man Rep must have both authority and responsibility, i.e. must be a senior role within management, however not necessarily part of top management. Saying that, in smaller companies, it's not uncommon for the CEO to also be the Man Rep. As the company grows, this is then delegated within the senior management team. 

The Man Rep can, and often has, its own “job” within the organisation, often within the Quality and Regulatory team, however, the Man Rep can also hold a position with other senior responsibilities in the organisation, as long as there isn’t a conflict of interest between the positions. For example, where the Man Rep is also the Head of Commercial, there will need to be controls in place to ensure commercial targets do not impact regulatory performance and QMS integrity.

What about their qualifications? Under ISO 13485 there is no explicit definition when it comes to qualifications, so in short the answer is there aren’t any specific qualifications or training requirements. However, as said earlier, the Man Rep must be part of the management team, with authority and responsibility.

There can only be one Man Rep who must be formally nominated and documented in a letter of appointment,  signed by Top Management with a clear start date and end date  (it can also be indefinite). The letter must define the Man Rep responsibilities and authority and must be saved within the QMS records and available to auditors. Yes, it’s one of the frequent questions, typically during a Stage 1 ISO 13485 audit, a quick win if you get it right.

What if the Man rep goes on holiday, or is sick? Although only one person in an organisation can be the Man Rep, under §5.5.1 the Man Rep can delegate his responsibilities to others. Here comes the Designee to save the day, the Designee is someone within the organisation who is delegated by the Man Rep with authority to carry out Man Rep activities. In practice this is often explicitly stated in the responsibility  assignment matrix (RAM) and can include authoring of procedures, reviewing of procedures, preparing the management review meetings, amongst other things.

A Man Rep must be appointed at the early stages of your QMS establishment, and formally appointed prior to your Stage 1 ISO 13485 audit. And what about the PRRC?

What is the role of PRRC?

The Person Responsible for Regulatory Compliance, or PRCC, is a mandatory role in the EU. Introduced by EU MDR 2017/745 Art.15:

“Manufacturers shall have available within their organisation at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of medical devices.”

and EU IVDR 2017/746 Art.15:  

“Manufacturers shall have available within their organisation at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of in vitro diagnostic medical devices.”

The PRRC ensures regulatory compliance not just from the QMS point of view, but covering all the device lifecycle, from the technical documentation to the post-market surveillance (PMS).

Who can be appointed as the PRRC?

Unlike the Man Rep where there are no specific requirements for qualifications, for the PRRC this is explicitely stated in EU MDR 2017/745 and EU IVDR 2017/746 Article 15, as well as in MDCG 2019-07 Rev.1:

“Manufacturers shall have available within their organisation at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of medical devices/in vitro diagnostic medical devices. The requisite expertise shall be demonstrated by either of the following qualifications: 

(a) a diploma, certificate or other evidence of formal qualification, awarded on completion of a university degree or of a course of study recognised as equivalent by the Member State concerned, in law, medicine, pharmacy, engineering or another relevant scientific discipline, and at least one year of professional experience in regulatory affairs or in quality management systems relating to medical devices/in vitro diagnostic medical device. 

(b) four years of professional experience in regulatory affairs or in quality management systems relating to medical devices/in vitro diagnostic medical devices. 

Without prejudice to national provisions regarding professional qualifications, manufacturers of custom-made devices may demonstrate the requisite expertise referred to in the first subparagraph by having at least two years of professional experience within a relevant field of manufacturing.”

In summary, this means having a relevant degree with at least 1 year of experience in regulatory affairs or quality management, OR 4 years of experience in regulatory or quality management. As with the Man Rep, the PRRC must be formally nominated in a letter of appointment, and the PRRC's qualifications and experience must be evidenced in the PRRC Curriculum Vitae, with reference to the appointment letter (again, this is often checked by the Notified Body during audits). 

There is nothing that precludes the Man Rep and the PRRC from being the same individual, provided there is no conflict in the roles, and of course they have the capacity to fulfil both roles. In addition, to ensure independence and avoidance of potential conflicts whilst performing their duties, Article 15(5) in both EU MDR and EU IVDR states:

“The person responsible for regulatory compliance shall suffer no disadvantage within the manufacturer’s organisation in relation to the proper fulfilment of his or her duties, regardless of whether or not they are employees of the organisation”

So it is implicit that the targets of the PRRC, i.e. to ensure that any product shipped satisfies the golden triad of being safe, effective and cybersecure, may be in conflict with targets of a delivery lead who may have targets relating to shipping to a timeline or a cost budget with less regard to the golden triad.  Whether the PRRC is internal or outsourced, an escalation route to Top Management, that is independent of the product delivery organisation, needs to be documented. 

As for the geographical location of the PRRC, as per MDCG 2019-7 it is recommended that it should be in proximity to the manufacturing activities for both within the organisation or externally outsourced.

When does a PRRC need to be in place? In the early stages of your QMS establishment, while in the pre-market stage, when there are no devices in the market, there is no legal requirement to appoint or outsource a PRRC. However, early regulatory input can save you time and money down the line, so engage with an external consultant, such as us folks here at Hardian Health to support you in the regulatory route.

A PRRC, however, must be formally in place, before you submit your technical documentation to the Notified Body, issue your Declaration of Conformity and put your CE mark in your device. 

For micro and small companies, though, it is not necessary to have a PRRC within the organisation; this role can be outsourced, as long as it can be permanently and continuously at your disposal. Hardian Health can act as your PRRC with a main PRRC named individual, as well as a deputy, ensuring, in the case of Article 15 EU (MDR 2017/745):

“The Person responsible for regulatory compliance shall at least be responsible for ensuring that:

(a) the conformity of the devices is appropriately checked, in accordance with the quality management system under which the devices are manufactured, before a device is released;

(b) the technical documentation and the EU declaration of conformity are drawn up and kept up-to-date;

(c) the post-market surveillance obligations are complied with in accordance with Article 10(10);

(d) the reporting obligations referred to in Articles 87 to 91 are fulfilled;

(e) in the case of investigational devices, the statement referred to in Section 4.1 of Chapter II of Annex XV is issued.”

For additional information on the PRRC services offered by Hardian Health, versus the manufacturer’s obligations, please see the tables below:

a)Checking device conformity before release

Article 15(2) of the EU Medical Device Regulation 2017/745 (MDR)

The PRRC service includes, in compliance with Article 15(3) of the MDR, the following items:

a) the conformity of the devices is appropriately checked, in accordance with the quality management system under which the devices are manufactured, before a device is released;

Article 15(2) of the EU In-vitro Diagnostic Medical Device Regulation 2017/746 (IVDR)

The PRRC service includes, in compliance with Article 15(3) of the IVDR, the following items:

a) the conformity of the devices is appropriately checked, in accordance with the quality management system under which the devices are manufactured, before a device is released;

Manufacturer is required to fulfil the following obligations

A. Change management
When a change to the QMS is proposed, the PRRC is offered the opportunity to review the proposed change request (the Change Control Record in the Hardian Core eQMS) and confirm whether that change is substantial or non-substantial before

  1. The Notified Body is notified of the change, per the terms of service of the Client’s Notified Body, to any administrative process (forms supplied by the Notified Body for the Client/ PRRC to fill, means of transmission e.g. email to a specified address, timeline for notification to the Notified Body and for response from the Notified Body)
    1. The change is enacted via the Document Management Procedure’s QMS change control provisions
    2. Changes are added to the agenda for Management Review and/ or the internal Audit Programme Plan as appropriate
  2. When a change to the product range covered by the CE certificate, or to any device, is proposed, the PRRC is offered the opportunity to review the proposed change request (the Change Control Record in the Hardian Core eQMS) and confirm whether that change is substantial or non-substantial before
    1. The Notified Body is notified of the change, per the terms of service of the Client’s Notified Body, to any administrative process (forms supplied by the Notified Body for the Client/ PRRC to fill, means of transmission e.g. email to a specified address, timeline for notification to the Notified Body and for response from the Notified Body)
    2. The change is enacted via the Design Control Procedure’s design change control provisions
    3. The PRRC is then responsible for sign-off on labelling in the release process, and as one of the signatories of the final Design Review before the changed device is put on market or into service, e.g.
      1. Software as a Medical Device (SaMD)
        1. user interface “about box” or suchlike having the correct version ID, including the correct UDI-DI and UDI-PI, or for software without a user interface, the equivalent labelling exposed via the Application Programming Interface (API)
        2. Any additional hardware labelling
        3. Changes to any Instructions for Use (IFU) and/ or other accompanying documentation
      2. Any additional hardware labelling
  3. In all cases the Management Representative is responsible for the timely communication with the Notified Body.
b)Technical documentation and the EU Declaration of Conformity

Article 15(2) of the EU Medical Device Regulation 2017/745 (MDR)

The PRRC service includes, in compliance with Article 15(3) of the MDR, the following items:

b) the technical documentation and the EU declaration of conformity are drawn up and kept up-to-date;

Article 15(2) of the EU In-vitro Diagnostic Medical Device Regulation 2017/746 (IVDR)

The PRRC service includes, in compliance with Article 15(3) of the IVDR, the following items:

b) the technical documentation and the EU declaration of conformity are drawn up and kept up-to-date;

Manufacturer is required to fulfil the following obligations

B. Same as A(b)(ii) in section a) above, with changes enacted via the Design Control Procedure’s design change control provisions.

c)Post-market surveillance

Article 15(2) of the EU Medical Device Regulation 2017/745 (MDR)

The PRRC service includes, in compliance with Article 15(3) of the MDR, the following items:

c) the post-market surveillance obligations are complied with in accordance with Article 10(10);

Article 15(2) of the EU In-vitro Diagnostic Medical Device Regulation 2017/746 (IVDR)

The PRRC service includes, in compliance with Article 15(3) of the IVDR, the following items:

c) the post-market surveillance obligations are complied with in accordance with Article 10(9);

Manufacturer is required to fulfil the following obligations

C. The relevant parties given in the PMS Plan are responsible to ensure that the PMS Report, PMCF/ PMPF Report, any Periodic Safety Update Report (PSUR) or MDR Summary of Safety and Clinical Performance (SSCP)/ IVDR Summary of Safety and Performance (SSP) are drafted and submitted to the PRRC for review at least two weeks prior to the due date for the respective document being completed, approved and where applicable submitted to the Notified Body.

d)Vigilance and incident reporting

Article 15(2) of the EU Medical Device Regulation 2017/745 (MDR)

The PRRC service includes, in compliance with Article 15(3) of the MDR, the following items:

d) the reporting obligations referred to in Articles 87 to 91 are fulfilled;

Article 15(2) of the EU In-vitro Diagnostic Medical Device Regulation 2017/746 (IVDR)

The PRRC service includes, in compliance with Article 15(3) of the IVDR, the following items:

d) the reporting obligations referred to in Articles 82 to 86 are fulfilled;

Manufacturer is required to fulfil the following obligations

D. The Management Representative or PRRC is responsible to file on EUDAMED and to submit to the Notified Body after completion of c).

e)Investigational and performance study devices

Article 15(2) of the EU Medical Device Regulation 2017/745 (MDR)

The PRRC service includes, in compliance with Article 15(3) of the MDR, the following items:

e) in the case of investigational devices, the statement referred to in Section 4.1 of Chapter II of Annex XV is issued.

Article 15(2) of the EU In-vitro Diagnostic Medical Device Regulation 2017/746 (IVDR)

The PRRC service includes, in compliance with Article 15(3) of the IVDR, the following items:

e) in the case of devices for performance studies intended to be used in the context of interventional clinical performance studies or other performance studies involving risks for the subjects, the statement referred to in Section 4.1 of Annex XIV is issued.

Manufacturer is required to fulfil the following obligations

E. The relevant parties given in the Clinical Evaluation Plan (CEP) or IVD Performance Evaluation Plan (PEP) are responsible to submit to the PRRC for review and sign-off any IVD Performance Evaluation Statement (PES) and per a) ii) 3) above any labelling marking the device as “NOT FOR CLINICAL USE” or “FOR PERFORMANCE EVALUATION ONLY” or “FOR RESEARCH USE ONLY” at least two weeks prior to any respective performance evaluation or clinical investigation commencing.

Do I need to register the PRRC? As of the 28th of May 2026 the PRRC must be registered as part of the manufacturer Actor Registration in EUDAMED, for both EU manufacturers and non-EU manufacturers as per EU MDR 2017/745 Art 31 (1) and EU IVDR 2017/746 Art 28.

EU manufacturers submit directly to EUDAMED as part of the initial Actor Registration (AR), note that if more than one PRRC, all individuals must be listed.

For non-EU manufacturers, the EU Authorised Representative (AR) must be appointed first,  and the AR needs to be registered in EUDAMED. The non-EU manufacturer must register directly with EUDAMED to start the Actor Registration process, which will include adding the PRRC details. The AR will then verify the information submitted and route for Competent Authority. 

An additional note for non-EU manufacturers, your AR must also have their own PRRC, which is for the AR to register when they register with EUDAMED. This also means that your PRRC as the manufacturer cannot be the same as the PRRC of your AR, again stated in the MDCG 2019-7. Do bear in mind that updates to the PRRC details must be done within 1 week - as per Art 31(4) for EU MDR and Art 28 (4) EU IVDR.

Hardian Health supports software and AI medical device manufacturers with QMS set-up, regulatory strategy and outsourced PRRC services, with a named PRRC and a deputy. If you're preparing for your Stage 1 audit or CE marking, our team can help.

Ana Ribeiro

Ana Ribeiro, Associate - Quality Assurance & Regulatory Affairs